FLIPR Calcium Assay Kit
Troubleshooting Guide
1. Fluorescence drop upon compound addition
This may be the result of dislodging cells from the wells during addition. Lowering the addition speed to 10-60
µl/sec should solve the problem in this case. If a lower addition speed eliminates the drop in fluorescence,
then seed fewer cells in the wells to avoid this problem in future.
If a fluorescence drop is still present after lowering addition speed, then it may be the result of a dilution
effect. Using the Loading Buffer at a higher concentration (as in the parameters in the table below) has been
shown to work in this case.
|
 |
| Parameters |
96-well Plate |
384-well Plate |
Dilution volume for loading buffer |
50 ml |
66 ml |
Addition volume for loading buffer |
50 µl |
25 µl |
Compound concentration (fold) |
4X |
4X |
Addition volume for compound (total for agonist and antagonist) |
50 µl |
17 µl |
2. Different compound concentration
The concentration of the Loading Buffer may be adjusted to compensate for compounds that are at different concentrations
from that recommended in the table in Section 3.2 of the Calcium Kit
Assay Protocol. The following table is provided as a guideline.
Note: It is not recommended to dilute the Loading Buffer to less than 33 ml. The Loading Buffer may become too
viscous for proper handling.
| |
Compound |
Loading Buffer |
| Plate |
Addition Volume (Agonist and Antagonist) |
Concentration (Fold) |
Dilution Volume |
Addition Volume |
96-well |
67 µl |
3X |
33 ml |
33 µl |
50 µl |
4X |
50 ml |
50 µl |
384-well |
17 µl |
4X |
66 ml |
25 µl |
25 µl |
3X |
66 ml |
25 µl |
3. Serum-sensitive cells or targets
Some cells are serum-sensitive, resulting in oscillations of intracellular calcium that could interfere with results.
Also, some target receptors or test compounds may interact with serum factors. In these cases, the growth medium
should be removed (and the cells may also need to be washed) prior to addition of Loading Buffer. The volume of
growth medium removed should be replaced with an equal volume of 1X Reagent Buffer before loading. Alternatively,
the cells could be incubated overnight in lower concentrations of FBS (i.e., 0.5%) and not washed prior to
the addition of Loading Buffer.
4. Cells tested with buffer plus DMSO show a calcium response
The buffer used for the negative control wells should contain the same final concentration of DMSO as is present in
the wells containing the test compounds. However, this concentration of DMSO could cause a calcium flux. In these
cases, add DMSO to the Loading Buffer so that the final concentration of DMSO in the wells will not change after
buffer addition.
5. Precipitation with the Loading Buffer
The FLIPR Calcium Assay Kit is compatible with numerous buffers. If the Loading Buffer prepared with 1X Reagent
Buffer causes precipitation (or some other unexpected effect), prepare the Loading Buffer with a different buffer
that has been shown to work in previously established assays.
Product Use Limitations and Warranty
All Molecular Devices reagent products are sold for research use only. Reagents may contain chemicals that are harmful.
Due care should be exercised to prevent direct human contact with the reagent.
Each product is shipped with documentation stating specifications and other technical information. Molecular
Devices products are warranted to meet or exceed the stated specifications. Molecular Devices' sole obligation and
the customer's sole remedy is limited to replacement of the products free of charge in the event that the product
fails to perform as warranted.
Molecular Devices Corporation makes no other warranties, either expressed or implied, including without limitation
the implied warranties of merchantability and fitness for a particular purpose or use.
|