SpectraTest
Fluorescence and Absorbance Reader Validation
applications
- Absorbance validation
- Fluorescence validation
- Automated testing
- NIST traceability
- Recertification
SpectraMax® absorbance and multi-detection microplate readers and Gemini fluorescence spectrophotometers are
designed to provide consistent performance for many years, but to fulfill your regulatory requirements, instrument
performance should be validated and documented regularly. The SpectraTest FL1 and SpectraTest ABS1 validation
packages provide automated, comprehensive and traceable validation of optical performance, and automatic
verification of your SpectraMax, VersaMax, Gemini EM and Gemini XPS.
SpectraTest FL1 fluorescence validation package tests
Qualify the fluorescence performance of the SpectraMax M2, SpectraMax M5, Gemini EM and Gemini XPS readers, by testing the specifications that are absolutely critical to compliance. Twelve different automated tests are provided:
- Fluorescence lower limit of detection (LLD)
- Excitation wavelength accuracy
- Emission wavelength accuracy
- Excitation wavelength precision
- Emission wavelength precision
- PMT matching (high vs. medium PMT settings)
|
- Top-to-bottom bias
- Kinetic noise (low signal)
- Kinetic noise (high signal)
- Well-to-well reproducibility
- RFU scale ratio
- Relative fluorescence unit (RFU) linearity
|
SpectraTest ABS1 absorbance validation package tests
Qualify the absorbance performance of the SpectraMax M2, SpectraMax M5, Plus384, Plus, 190, 340PC, 340PC384 and VersaMax
readers by testing the specifications that are absolutely critical to compliance. Eight different automated tests are
provided:
- Absorbance accuracy (linearity)
- Precision (reproducibility)
- Stray light
- Wavelength accuracy
|
- Wavelength repeatability
- Ultimate dark (0% Transmittance)
- Optical alignment
- Baseline noise
|
automated for ease of use
All test measurements and calculations are handled automatically by the software protocols. Should any of the
measurement parameters fall outside defined limits, a test failure is reported with the suspect parameters identified.
NIST traceability
SpectraTest ABS1
Each filter is individually tested across all eight channels of the plate. Calibration of the plate's filter standards
is accomplished through the use of an instrument calibrated with primary NIST standards.
SpectraTest FL1
A combination of fluorescence reference material and NIST-traceable didymium glass exhibits calibrated wavelength
peaks used for wavelength accuracy testing.
recertification service
So that confidence in the standards is maintained, Molecular Devices recommends having your validation plate recertified
at one year intervals. Validation plates sent to Molecular Devices for recertification are disassembled, cleaned and
calibrated, then returned with a new certificate of calibration.
additional validation tools
In addition to the SpectraTest ABS1 and FL1 validation plates, IQ/OQ/PQ validation, SoftMax® Pro software
validation and FDA 21 CFR Part 11 compliance tools are available.
|