Molecular Devices Home
Careers  |  Contact Us  |  Search:   

  

Automated Electrophysiology Comparison Chart

  IonWorks
Quattro
PatchXpress
7000A
OpusXpress
6000A
Applications Primary screening of directed libraries, secondary screening, lead optimization, early toxicology screens Compound profiling, lead optimization, structure-activity relationship studies, cardiac safety assessment lead optimization, basic research, target discovery
Electrode Type Planar Planar Pipette
Substrate Polyimide Glass Glass
Recording Chambers 384 16 8
Patching Technology Single-hole (SH) and Population Patch Clamp (PPC) Single-Hole Planar Patch Clamp Two-Electrode Voltage Clamp
Recording Configuration Perforated Patch Clamp Whole-Cell Patch Clamp Two-Electrode Voltage Clamp
Typical Success Rate

SH: 60% – 90%

PPC: >95%
50% – 90% 90% – 100%
Cell Type CHO, HEK, CHL, NIH3T3, BHK, L929, NG108, Rin5F, SF9, and others HEK, CHO, CHL, RBL, L(tk), SH-SY5Y, GH4C1, LNCaP, SJ-RH30, PC12, 3T3 and others Oocytes
Compatible Ion-
Channel Types
Voltage-Gated and Slowly Desensitizing/Activating Ligand-Gated Ion Channels Voltage-Gated and Rapidly Desensitizing Ligand-Gated Voltage- and Ligand-Gated Ion Channels
Seal Resistance 0.05 - 1.0 GΩ 1.0 - 3.0 GΩ Not Applicable
Membrane Resistance (Rm) Not Measured 0.2 - 2 GΩ 0.5 - 10 MΩ
Data Points/6 hr Experimental Time 2300 250 60
Access Resistance (Rs) 10 - 15 MΩ 3 - 8 MΩ Not Measured
Rs Control None Pneumatic, Feedback Controlled Not Applicable
Rs Compensation None >=80% Not Applicable
Membrane Capacitance (Cm) Compensation No Yes No
Pipette Tips Fixed Disposable Disposable
Number of Pipettes 12 or 48 1 8
Recording During Compound Addition No Yes Yes
Time Between Compound Addition and Recording 7 s Minimum Continuous Add and Read Continuous Add and Read
Compound Addition Time Constant Not Applicable tau<15 ms 1-3 s for 90% exchange
Compound Wash Out No Yes Yes
Compound Wash Out Volume Not Applicable 1000 mL total User-Configurable
System Control Software IonWorks Quattro PatchXpress Commander OpusXpress Commander
Data Acquisition Rate 156 Hz – 10 kHz 1 Hz – 31.25 kHz 1 Hz – 31.25 kHz
Real Time Measurements Display Yes Yes Yes
Control of Patching Parameters Yes Yes Yes
Data Management No DataXpress 2.0 No
Data Analysis No DataXpress 2.0 No
Automated Data Analysis No Yes No