MultiClamp 700B |
 |
Computer-Controlled Dual Channel Resistive-Feedback Patch Clamp and High-Speed Current Clamp Amplifier
The MultiClamp 700B is useful for a wide variety of intracellular or extracellular
recording, including:
- High-speed Current Clamp (sharp-electrode or field potentials)
- Patch Clamp (whole-cell, macro-patch or excised patch)
- Voltammetry / Amperometry
- Ion-selective measurement
- Bilayer recording
For complete information and specifications, download the MultiClamp 700B manual from the
Literature link in the right menu.
Major New Features
- Internal automatic mode switch
Instantly switch recording modes by using the membrane potential as the trigger. Pre-set a voltage threshold
in current-clamp mode, and MultiClamp 700B will automatically switch to voltage-clamp when the membrane
potential reaches that voltage level. User-specified delay will allow you to perform creative new experiments.
- Oscillation suppression
Sudden changes in membrane or pipette parameters may result in undesirable oscillations during whole cell
recording. MultiClamp 700B will detect current or voltage oscillations and automatically disable or
intelligently reduce compensation settings to protect your cell from damage.
- Slow current injection
Slight voltage drift—often due to changing electrode properties—may contaminate an otherwise
decent current-clamp recording. In order to maintain the membrane potential at a consistent level,
MultiClamp 700B will automatically inject compensatory current over a user-defined time course.
Software Control

Instead of the usual front panel knobs and switches, the MultiClamp 700B is controlled by the
MultiClamp 700B Commander, a program that runs on a host computer (PC and Macintosh; see Requirements below)
and communicates with the amplifier via a USB cable. This control interface provides automation of bridge balance,
pipette offset, pipette and whole-cell capacitance compensation, in addition to "smart" features such as protection
from oscillations and threshold-based mode change. Amplifier settings such as gain, filter frequency,
whole-cell capacitance, recording mode and also input / output scale factors are automatically telegraphed
to the data acquisition software via "messaging" through the computer operating system.
For a more in-depth look at the program, you can download the MultiClamp 700B Commander to try on your
Windows computer by going to the
Literature page.
Optional SoftPanel

Although the MultiClamp 700B is a computer-controlled amplifier, the mouse and/or keyboard is not the only
means of controlling the instrument. The optional SoftPanel was designed for those who prefer a more
conventional feel to amplifier control. By way of a USB connection, the SoftPanel physically replicates all
essential amplifier functions by acting as a hardware extension of the MultiClamp 700B Commander software.
SoftPanel knobs replicate continuous mouse controls ("gliders"), while buttons replicate single-click mouse controls.
Headstages

|
The MultiClamp 700B comes standard with two identical but independent CV-7B headstages, each of which
contains both current-to-voltage and voltage-following circuitry. This design allows the user to rapidly
switch between patch clamp recording and true high-speed current clamp recording. Thus, with two headstages,
the MultiClamp 700B can perform the function of two patch clamps, two current clamps, or a combination of
patch and current clamp amplifiers. Furthermore, two optional voltage-follower headstages (HS-2 type) can
be connected to auxiliary inputs to allow third and fourth-point voltage recording.
|
|
MultiClamp 700B supports up to four simultaneous headstages. Two CV-7B headstages come standard;
the HS-2A headstages are optional.
|
The CV-7B was designed for low noise and flexible recording features. However, some specialized applications
require even more flexibility. For example, the rather large membrane capacitance in bilayer recording demands
greater capacitance compensation. The optional CV-7B/BL headstage was developed for this purpose. Another
optional headstage, the CV-7B/EC, was designed for large (± 2 V) commands required during electrochemistry
recording (amperometry, voltammetry).
Requirements

Computer Control
The MultiClamp 700B Commander program runs on the US Windows Vista/XP/2000/ME/98 operating systems, as well as
Macintosh OS X operating system, versions 10.4.6 - 10.4.11 (OS 9 is not supported) on an Intel or Power PC-based computer.
A USB port is required to connect the MultiClamp 700B. An additional USB port is required if the optional
SoftPanel is used to control the amplifier.
Experimental Control and Data Acquisition
Although the MultiClamp 700B is controlled by a software interface, it remains a conventional analog input/output
amplifier. Thus, it requires a system for controlling your stimulus protocols and recording the output.
Our pCLAMP software and Digidata 1440A
digitizer provide the most complete, integrated solution. Note: pCLAMP 9 or
higher is required for automatic telegraphing.
Ordering Information

List of technical arguments for sole source justification is available on the
Literature page.
MultiClamp 700B
- Comes standard with two CV-7B headstages.
|