SpectraMax 340PC384
Microplate Spectrophotometer
Features
- No filters needed.
- Easy assay optimization.
- Well volume sensor.
- Comprehensive data analysis software.
- Optical validation.
The SpectraMax 340PC384 provides everything you need to measure absorbance in the visible range,
including temperature control, robotics-compatible interface and
SoftMax® Pro data analysis software.
accurate quantitation
The optical design of the SpectraMax 340PC384 truly mimics a dual-beam spectrophotometer. Each well has
its own sample beam and reference detector. The eight-channel system, consisting of eight sample beams and eight
reference beams, delivers both superior precision and speed of reading across the microplate.
no interference filters required
The SpectraMax 340PC384 has a monochromator instead of interference filters. Simply select the absorbance
maximum of your sample and let the monochromator tune to that exact wavelength. Or scan up to 384 wells at any time
to determine the best assay conditions. The wavelength and scanning ranges are from 340-850 nm in 1 nm increments.
It is like having 511 filters built-in so you will always have the correct wavelength.
measure fluid volumes in individual wells
The patented* PathCheck® sensor measures the depth of the liquid in each microplate well. This information can
be used to normalize absorbance values to a 1 cm pathlength so they will agree with measurements made in a standard
1 cm cuvette. You can also use this feature to calculate concentrations without a standard curve and expand the
dynamic range to 6+ OD. PathCheck can accurately measure liquid volumes down to 5 µl to test the accuracy and
precision of liquid handling devices.
applications
- ELISA/EIA
- Cytoproliferation/Cytotoxicity
- Kinetic ELISAs/Enzyme Assays
- β-galactosidase Reporter Gene
- Colorimetric Protein
- NAD(P)H-based assay
leading microplate data analysis software
SoftMax® Pro microplate data analysis software provides flexibility in experimental design, set-up, analysis and
reporting, giving you the opportunity to customize your assays exactly as you require. You can password-protect your
data, choose from nine different curve-fitting routines and use default data reduction, or set up your own formulas
for analysis. You can even analyze and combine data from different plates. (See Figures 1, 2 and 3.)
 |
 |
 |
|
 |
 |
Figure 1. Standards for multiple calibration curves and unknowns can be run on separate plates. |
 |
 |
 |
|
 |
|
 |
 |
|
 |
|
 |
 |
|
 |
 |
 |
 |
 |
|
 |
 |
Figure 2. Multiple calibration curves can be plotted on one graph. |
 |
 |
 |
|
 |
|
 |
 |
|
 |
|
 |
 |
|
 |
 |
 |
 |
 |
|
 |
 |
Figure 3. Results from unknowns run on different plates and different calibration curves can be plotted on one graph. |
 |
 |
 |
|
 |
|
 |
 |
|
|
|
 |
 |
|
 |
 |
optical performance validation
The optional SPECTRAtest™ Validation Package provides an NIST-traceable solution
for validating the SpectraMax 340PC384. The following measurements are automatically made by SoftMax Pro:
baseline noise, dark noise, optical alignment, wavelength accuracy, photometric linearity and wavelength precision.
Specifications
photometric performance |
Wavelength Range |
340-850 nm |
Photometric Range |
0 to 4.000 OD |
Wavelength Selection |
Monochromator, tunable in 1-nm increments |
Photometric Linearity (405 nm) |
0 to 3.000 OD |
Wavelength Bandwidth |
2 nm |
Photometric Accuracy (340-850 nm) |
< or = 0.006 OD ± 1.0%, 0-2 OD |
Wavelength Accuracy |
<± 1.0 nm |
Photometric Precision |
< or = 0.003 OD ± 1.0%, 0-2 OD |
Wavelength Repeatability |
± 0.2 nm |
PathCheck Measurement Error |
< 5% compared to a 1 cm cuvette |
microplate read time (single wavelength) |
Endpoint Read Time |
96 wells in 9 seconds 384 wells in 29 seconds |
Minimum Kinetic Interval |
96 wells in 9 seconds 384 wells in 29 seconds |
temperature regulation |
Temperature Range |
Ambient + 4 °C to 45 °C |
* US Patent Nos. 5,959,738 6,188,476 6,320,662 6,339,472 6,404,501, 6,496,260.
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
|