SpectraMax 190
Microplate Reader with PatchCheck®
Features
- Use the extinction coefficient of your sample to calculate concentrations directly.
- Correct for volume differences in the wells.
- Expand the dynamic range of your assay.
- Detect pipetting errors.
- Check performance of your pipetting devices.
The SpectraMax® 190 spectrophotometer is ideal for most life science applications, especially DNA analysis.
For aqueous solutions, our patented* PathCheck® sensor technology senses the depth of the liquid
in each microplate well, and normalizes the absorbance value to a 1-cm pathlength. The corrected absorbance
is within 5% of the value obtained in a conventional spectrophotometer.
Industry leading SoftMax® Pro microplate analysis software provides comprehensive
instrument control and data reduction.
spectral scanning makes assay development easy
You can select the optimum wavelength for reading your assay by viewing the complete spectrum of your sample, allowing
you to monitor the stability of your sample or reagent, determine interfering compounds or conditions of your assay,
and validate your assay to regulatory standards.
accurate nucleic acid measurements
The SpectraMax 190 microplate spectrophotometer is the tool you need for DNA analysis. A bandwidth of 5 nm or less
is required for accurate absorbance measurement of DNA1. With the ability to detect DNA down to 25 ng per
well, as well as a limit of calculation to 75 ng per well, the SpectraMax 190 microplate spectrophotometer gives you
the same sensitivity as a conventional spectrophotometer.
expand the dynamic range of any assay to 0 to 6+ OD
The PathCheck sensor can extend the instrument linearity by a factor of three or more. For very concentrated samples,
use a minimum volume in the microplate well (100 µl has a pathlength of approximately 0.30 cm), then use the
PathCheck sensor to calculate an equivalent 1-cm pathlength absorbance value, giving you calculated absorbance values
well beyond the capabilities of other instruments.
quickly determine multi-channel pipettor performance
Precision and accuracy of multi-channel pipettors can now be verified using the actual dispense reagent without the
need to add a dye. By measuring the depth of the liquid in the well, the patented PathCheck sensor can determine the
volume of the liquid dispensed2. Even 96-channel dispensers can be tested in a matter of minutes.
early ADME compound profiling assays
SpectraMax microplate readers are ideal for increased throughput of early ADME compound profiling assays such as the
Transil® Membrane Affinity Assay and Human Serum Albumin Binding
Assay. The Membrane Affinity assay provides high-throughput lipophilicity data equivalent to liposome-based methods
without the need for ultracentrifugation. The Human Serum Albumin Binding assay provides high-throughput blood protein
binding data by measuring the fraction bound to Human Serum Albumin.
Specifications
photometric performance
|
Wavelength Range |
190 - 850 nm |
Wavelength Selection |
Monochromator, tunable 1.0 nm increments |
Wavelength Bandwidth |
2 nm |
Wavelength Accuracy |
<± 2.0 nm |
Wavelength Repeatability |
± 0.2 nm |
Photometric Range |
-0.3 to 4.0 OD |
Photometric Resolution |
0.001 OD |
Photometric Accuracy (190-850 nm) |
< ±0.006 OD ±1.0%, 0 - 2.0 OD |
Photometric Precision (190-850 nm) |
< ±0.003 OD ±1.0%, 0 - 2.0 OD |
PathCheck Measurement Error |
< 5% when compared to cuvette reading |
Stray Light |
< 0.05% @ 230 nm |
Light Source |
Xenon flash lamp |
Microplate Read Time: Endpoint Kinetics |
12 seconds 9 second minimum interval |
Temperature Regulation Specifications |
Temperature Range |
Ambient + 4 °C up to 45 °C |
Temperature Uniformity (Microplate) |
± 0.5 °C @ 37 °C, well-to-well |
Temperature Stabilization Time |
30 minutes maximum upon initiation |
Reading Chambers |
Isothermal when temperature regulation not enabled, <1 °C |
physical specifications |
Dimensions |
22 cm (h) x 58 cm (w) x 38 cm (d) |
Weight |
30 lb (13.6 kg) |
* US Patent Nos. 5,959,738 6,188,476 6,320,662 6,339,472 6,404,501, 6,496,260.
|