Home Page | Products | Instruments | Microplate Readers | Multi Mode Readers | SpectraMax Paradigm | LUM Cartridge
SpectraMax Paradigm Luminescence Detection Cartridge
The Luminescence Detection Cartridge utilizes a unique, well-isolating design that significantly reduces the level of cross talk from surrounding wells. The resulting performance is ideal for sensitive chemiluminescence applications.
Key Applications
- Apoptosis
- cAMP Quantitation
- GPCR Ligand Binding
- GTP Binding
- Immunoasssay
Technical Specifications
| Luminescence (Lum) | Guaranteed* | Optimized |
|
96-well (200 µL) |
0.50 fmol | 0.30 fmol |
| 384-well (75 µL) | 0.50 fmol | 0.30 fmol |
| 1536-well (8 µL) | 0.20 fmol | 0.15 fmol |
|
|
||
|
96-well (200 µL) |
0.40 fmol | 0.25 fmol |
| 384-well (75 µL) | 0.20 fmol | 0.15 fmol |
|
|
||
|
96-well (200 µL) |
0.30 fmol | 0.20 fmol |
* For properly functioning, operated and maintained equipment.
Cartridge Details
| Part No. | 0200-7014, 0200-7015, 0200-7016 |
| Mounting | Top or bottom read |
| Cartridge Size | 1 position |
| Emission Range | Visible to 650 nm |
Related Information
- TUNE Cartridge (PDF)
- Multi-Mode Detection Cartridge (PDF)
- ABS Cartridge (PDF)
- AlphaScreen Cartridge (PDF)
- FI Cartridge (PDF)
- FI Dual-Label Cartridge (PDF)
- FP Cartridge (PDF)
- FI GeneBLAzer Cartridge (PDF)
- HTRF Cartridge (PDF)
- LUM Cartridge (PDF)
- Dual-Color LUM Cartridge (PDF)
- MULTI Cartridge (PDF)
- TRF Cartridge (PDF)
- CatchPoint cAMP Detection Assay Kit
- CatchPoint cGMP Detection Assay Kit
- IMAP Kinase, Phosphatase, and Phosphodiesterase Assays
- Molecular Devices Releases Breakthrough TUNE Technology, Now Part of the SpectraMax Paradigm Multi-Mode Microplate Detection Platform
What's New?

Read What You Want When You Want with the new user-upgradeable SpectraMax Paradigm Multi-Mode Microplate Reader. Learn more.